  Datasheet File OCR Text:
 1~4 Output Medical Type
120W Medical
Features
Universal AC input / Full range Low leakage current <180 A Protections: Short circuit / Overload / Over voltage UL60601-1 medical safety approved With power good and fail signal output (Optional) 100% full load burn-in test Fixed switching frequency at 45KHz 3 years warranty
series
EN60601-1 IEC60601-1
Specification
AC INPUT VOLTAGE 90~264 VAC, 47~440Hz / 127~370VDC. AC INPUT CURRENT (Typ.) Maximum input current 2.9A at 115VAC, 60Hz or 1.7A at 230VAC, 60Hz with 100% output load. INRUSH CURRENT (Typ.) Inrush current is less than 22A at 115VAC or less than 45A at 230VAC under cold start conditions. Limiting provided by internal thermistors. SETUP, RISE TIME MPS-120: 800ms, 20ms / 230VAC at full load 2000ms, 50ms / 115VAC at full load MPD,T,Q-120:500ms, 20ms / 230VAC at full load 1200ms, 50ms / 115VAC at full load HOLD-UP TIME (Typ.) 80ms / 230VAC at full load 14ms / 115VAC at full load LEAKAGE CURRENT Leakage current is less than 180 A at 264VAC DC OUTPUT ADJ. RANGE DC output voltage (or Ch1 of multiple output models) can be adjusted between -5%~+10% rated output voltage by potential meter. OVERLOAD PROTECTION Fully protected against short circuit and output overload. The hiccup type protection will be activated at 110~150% (For MPD,T,Q-120),120~160%(For MPS-120) rated load and recovers automatically after fault condition is removed. OVER VOLTAGE PROTECTION Provided on output channel 1 only at 115%~135% rated output voltage. (120%~140% for MPS-120-15/24/48). Output will be shut down when this protection is activated. POWER GOOD / FAIL SIGNAL (OPTIONAL) TTL logic high for power good and TTL low for power fail. When the output voltage reaches 90% of rated value, a +5V TTL signal will be sent out with a 10~500ms delay; At least 1ms before the output voltage goes below 90% of the rated value, the TTL signal will be turned off. * MPS-120-3.3 does not have this optional function. WORKING TEMP. Whole series can operate from -20~70 to the derating curves. WORKING HUMIDITY 20~90% RH non-condensing. STORAGE TEMP., HUMIDITY -40~+85 , 10~90% RH . Please refer
TEMP. COEFFICIENT 0.04%/ on all outputs at full load between 0~50 of ambient temperature.
VIBRATION 2G of acceleration, vibrating frequency adjust from 10Hz ~500Hz within a 10-minute cycle. 6 testing cycles (60 minutes) each along X, Y, Z axes. SAFETY STANDARDS Medical : UL60601-1, TUV EN60601-1, IEC60601-1 approved Commercial : Also design refer to UL60950-1, TUV EN60950-1 WITHSTAND VOLTAGE 4000VAC between input and output 1500VAC between input and F.G. 500VAC between output and F.G. ISOLATION RESISTANCE >100M Ohms for I/P-O/P, I/P-FG, O/P-FG by using 500VDC test voltage. EMI COMPLIANCE EMI Specifications Conducted & Radiation Harmonic distortion Voltage flicker EMS COMPLIANCE EMS Specification ESD air ESD contact RF field susceptibility EFT(Electrical Fast Transient)/Burst Lightning/Surge Conducted RF susceptibility Magnetic field immunity Voltage dip, interruption Digital phone carrier immunity MTBF 262,100 hours min. at full load and 25 calculated per MIL-HDBK-217F. DIMENSION (L*W*H) 177.8x107.95x35.5mm or 7"x4.25"x1.4" PACKING 0.55Kg; 24pcs/14.5Kg/1.04CUFT Compliance Level EN55011, Class B EN55022, Class B EN61000-3-2 EN61000-3-3 Compliance Level EN61000-4-2, Level 3, 8KV EN61000-4-2, Level 2, 4KV EN61000-4-3, Level 2, 3V/m Level 3, 10V/m EN61000-4-4, Level 2, 1KV/5KHz Level 3, 2KV/5KHz EN61000-4-5, Level 4, 2KV/Line-Line 4KV/Line-Earth EN61000-4-6, Level 2, 3Vrms/m Level 3, 10Vrms/m EN61000-4-8, Level 2, 3A/m Level 3, 10A/m EN61000-4-11,Compliance ENV50204, Level 2, 3V/m, 900MHz Level 3, 10A/m, 900MHz of ambient temperature,
File Name:120W MEDICAL-SPEC 2008-03-25
1~4 Output Medical Type
Output Chart
MODEL MPS-120-3.3 MPS-120-5 MPS-120-12 MPS-120-15 MPS-120-24 MPS-120-48 MPD-120A OUTPUT VOLTAGE 3.3V 5V 12V 15V 24V 48V 5V 12V MPD-120B 5V 24V MPT-120A 5V 12V -5V MPT-120B 5V 12V -12V MPT-120C 5V 15V -15V MPT-120D 5V 24V 12V MPQ-120B 5V 12V -5V -12V MPQ-120C 5V 15V -5V -15V MPQ-120D 5V 12V 24V -12V MPQ-120E 5V 12V 15V 24V
Notes :
120W Medical
WITH FAN (25CFM) 24A 22A 10A 8A 5A 2.5A 10A 5A 10A 2.9A 10A 4.8A 0.6A 10A 4.4A 0.6A 10A 4A 0.6A 10A 2.2A 0.6A 10A 4.2A 0.6A 0.6A 10A 3.2A 0.6A 0.6A 10A 1A 2.1A 0.6A 10A 3A 0.6A 0.6A RIPPLE & NOISE VOLTAGE PEAK LOAD (Max.) TOLERANCE WITH 25CFM FAN (Note 2) (Note 3) 26A 26A 11A 8.8A 5.5A 2.8A 12A 6A 12A 3.2A 12A 5.8A 1A 12A 5.3A 1A 11A 4.4A 1A 12A 2.64A 1A 11A 5A 1A 1A 11A 3.8A 1A 1A 11A 1.1A 2.3A 1A 11A 3.3A 1A 1A 80mVp-p 80mVp-p 100mVp-p 100mVp-p 120mVp-p 120mVp-p 80mVp-p 120mVp-p 80mVp-p 250mVp-p 80mVp-p 120mVp-p 80mVp-p 80mVp-p 120mVp-p 80mVp-p 80mVp-p 150mVp-p 80mVp-p 80mVp-p 300mVp-p 120mVp-p 80mVp-p 120mVp-p 80mVp-p 80mVp-p 80mVp-p 150mVp-p 80mVp-p 80mVp-p 80mVp-p 150mVp-p 300mVp-p 80mVp-p 80mVp-p 120mVp-p 80mVp-p 80mVp-p 3.0% 3.0% 2.0% 2.0% 2.0% 2.0% 2.0% 7.0% 2.0% 8.0% 2.0% +8,-6% 5.0% 2.0% 6.0% 5.0% 2.0% +6,-7% 5.0% 2.0% +8,-6% 5.0% 2.0% 6.0% 5.0% 5.0% 2.0% +6,-7% 5.0% 5.0% 2.0% +8,-6% 8.0% 5.0% 2.0% 6.0% 8.0% 5.0%
series
RATED CURRENT 24A 22A 10A 8A 5A 2.5A 10A 5A 10A 2.9A 10A 4.8A 0.6A 10A 4.4A 0.6A 10A 4A 0.6A 10A 2.2A 0.6A 10A 4.2A 0.6A 0.6A 10A 3.2A 0.6A 0.6A 10A 1A 2.1A 0.6A 10A 3A 0.6A 0.6A
OUTPUT CURRENT MINIMUM LOAD 0A 0A 0A 0A 0A 0A 2A 0.5A 2A 0.3A 2A 0.4A 0A 2A 0.4A 0A 2A 0.4A 0A 2A 0.4A 0A 2A 0.5A 0A 0A 2A 0.5A 0A 0A 2A 0.2A 0.3A 0A 2A 0.5A 0A 0A CONVECTION (max.) 16A 14.7A 6.7A 5.3A 3.3A 1.7A 7.3A 3.6A 7A 1.9A 7.3A 3.5A 0.3A 7.3A 3.2A 0.4A 7A 2.6A 0.4A 7.3A 1.6A 0.4A 7.3A 3.1A 0.4A 0.4A 7.3A 2.4A 0.4A 0.4A 7A 0.7A 1.4A 0.3A 7.3A 2.3A 0.4A 0.4A
LINE LOAD REGULATION REGULATION EFFICIENCY 1.0% 1.0% 1.0% 1.0% 1.0% 1.0% 0.5% 2.0% 0.5% 2.0% 0.5% 1.5% 0.5% 0.5% 1.5% 0.5% 0.5% 2.0% 0.5% 0.5% 3.0% 0.5% 0.5% 1.5% 0.5% 0.5% 0.5% 2.0% 0.5% 0.5% 0.5% 2.0% 2.0% 0.5% 0.5% 2.0% 2.0% 0.5% 3.0% 3.0% 2.0% 2.0% 2.0% 2.0% 0.5% 3.5% 0.5% 4.0% 1.0% 3.5% 1.0% 1.0% 3.5% 1.0% 1.0% 3.5% 1.0% 1.0% +4,-3% 1.0% 0.5% 3.5% 1.0% 1.0% 0.5% 3.5% 1.0% 1.0% 0.5% 3.5% 3.5% 1.0% 0.5% 3.0% 3.0% 1.0% 73% 74% 71% 71% 74% 72% 73% 72% 68% 73% 77% 79% 81% 82% 75%
(Note 4)
76%
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature. 2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1 f & 47 f parallel capacitor. 3. Tolerance : includes set up tolerance, line regulation and load regulation. 4. 33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power. 5. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives.
File Name:120W MEDICAL-SPEC 2008-03-25
1~4 Output Medical Type
Mechanical Specification
177.8[7"] 165.1 [6.5"]
120W Medical
series
Unit:mm
CN1 5 4 3 2 1 13 12 11 10 9 8 7 6 5 4 3 2 1 CN2
+V ADJ
10cm
4-
3.96[0.156"]
FAN
35.5 [1.4"]
2 [0.078"] max.
Pin Assignment
AC Input Connector (CN1) : Molex 5273-05 or equivalent
Pin No. 1 2,4 3 5 MPS-120 FG No Pin AC/N AC/L MPD-120 FG No Pin AC/N AC/L MPT-120 FG No Pin AC/N AC/L MPQ-120 FG No Pin AC/N AC/L Mating Housing Molex 5195 or equivalent Terminal Molex 5194 or equivalent
DC Output Connector (CN2) : Molex 5273-13 or equivalent
Pin No. 1 2 3 4 5 6 7 8 9 10 11 12 13 MPS-120 +V +V +V -V -V -V -V +V +V P.F.D. NC No Pin NC MPD-120 V1 V1 V1 COM COM COM COM V2 V2 P.F.D. NC No Pin NC MPT-120 V1 V1 V1 COM COM COM COM V2 V2 P.F.D. V3 No Pin NC MPQ-120 V1 V1 V1 COM COM COM COM V2 V2 P.F.D. V3 No Pin V4 Mating Housing Terminal
Molex 5195 or equivalent
Molex 5194 or equivalent
Derating Curve
100
Static Characteristics
100
25CFM forced air cooling
80
90
MPS/D-120 Convection cooling 53W MPS-120-3.3 only 73.7W MPS-120-5 only 80W others
80 70
60
LOAD (%)
40
MPT/Q-120
LOAD(%)
60 50 40
20
-20
0
10
20
30
40
50 55 60
70 (HORIZONTAL)
90
100
115
107.95[4.25"]
135
95.25[3.75"]
155
230 264
AMBIENT TEMPERATURE ( )
INPUT VOLTAGE (V) 60Hz
File Name:120W MEDICAL-SPEC 2008-03-25


▲Up To Search▲   

 
Price & Availability of MPT-120A

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X